Specifications Show all

Controls
IPAP
50 cm H₂O
CPAP
4 - 20 cmH20 (passive leak port circuits) cm H₂O
EPAP/PEEP (a)
0 - 25 cmH20 (active valve circuits) cm H₂O
EPAP/PEEP (b)
4 - 25 cmH20 (passive leak port circuits) cm H₂O
Pressure support
0 – 30 cmH2O cm H₂O
Tidal volume
50 - 2000 ml
Breath rate(a)
0 - 60 BPM (AC mode) beats per minute
Breath rate(b)
1 - 60 BPM (all other modes) beats per minute
Inspiratory time
0.3 - 5.0 s Synchrony features
Rise time
1 - 6 (relative scale)
Ramp start pressure (a)
0 - 25 cmH20 (active circuits) cm H₂O
Ramp start pressure (b)
4 - 25 cmH20 (passive circuits) cm H₂O
Ramp start pressure (c)
4 - 19 cmH20 (CPAP mode) cm H₂O
Ramp time Off
5 - 45 min
C-Flex Off
1 - 3 (relative scale)
Flow trigger sensitivity
1 - 9 l/min
Flow cycle sensitivity
10 - 90 %
Environmental
Operating temperature
5 – 40 °C
Relative humidity
15 - 95 %
Storage temperature
-20 – 60 °C
Atmospheric pressure
60 - 110 kPa (450 - 825 mmHg)
Patient types
Pediatric
≥ 5 kg or greater
Adult
Yes
Electrical
Input voltage
100 - 240 VAC and 50/60 Hz and 2.1 A
Detachable battery voltage
14.4 VDC
Internal battery life
3 hours under normal conditions hr
Detachable battery life
3 hours under normal conditions hr
external battery connection
12 VDC
Volume modes
Pressure control (PC)
Yes
MPV Mouthpiece Ventilation Mode
Yes
Synchronized intermittent mandatory ventilation (SIMV)
Yes
SIMV with pressure support (SIMV w/PS)
Yes
Control ventilation (CV)
Yes
Alarms
Circuit disconnect Off
10 - 60 sec
Low tidal volume Off
50 - 2000 ml
Apnea Off
10 - 60 s and 4 - 60 BPM
High tidal volume Off
50 - 2000 ml
High minute ventilation Off
1 - 99 l/min Off, 1 - 99 l/min
Low minute ventilation Off
1 - 99 l/min
High respiratory rate Off
4 - 80 beats per minute
Low respiratory rate Off
4 - 80 beats per minute
Pressure modes
Pressure control (PC)
Yes
Pressure control-SIMV (PC-SIMV)
Yes
Spontaneous ventilation (S)
Yes
Spontaneous ventilation with timed back-up (S/T)
Yes
Timed ventilation (T)
Yes
Continuous positive airway pressure (CPAP)
Yes
Average volume assured pressure support (AVAPS)
With passive circuit type, S, S/T, PC, and T modes only
Options
Detachable back-up battery
Offering up to 3 hours additional operating time
Hospital roll stand
Provides convenient accessory basket and humidifier mount
DirectView
1 GB SD card data storage integrated into the ventilator software
Monitored parameters
Tidal volume
0 - 2000 ml
Minute ventilation
0 - 99 l/min
Estimated leak rate
0 - 200 l/min
Respiratory rate
0 - 80 BPM 0 - 80 beats per minute
Peak inspiratory flow
0 - 200 l/min
Peak inspiratory pressure
0 - 99 cm H₂O
Mean airway pressure
0 - 99 cm H₂O
% patient triggered breaths
0 - 100 %
I:E ratio
9.9:1 - 1:9.9
Synchrony features
Auto-Trak sensitivity
Auto-Trak sensitivity Auto-adaptive triggering, cycling, and leak compensation (available in all modes, passive circuit only)
Adjustable flow triggering
1 – 9 l/min (available in all modes and circuit types) l/min
Circuit types
Active exhalation valve with proximal airway pressure (PAP)
Yes
Active exhalation valve with flow sensor
Yes
Passive exhalation port
Yes
Compliance
IEC 60601-1 Medical electrical equipment Part 1: General requirements for safety
Yes
IEC 60601-1-2 General requirements for safety – collateral standard Electromagnetic compatibility – requirements and tests
Yes
IEC 60601-2-12 Medical electrical equipment Part 2-12: Particular requirements for the safety of lung ventilators – Critical Care ventilators
Yes
Oxygen
FiO2
21 - 100 %
02 flush
2 min at 100%
02 input pressure range
276 - 600 kPa (40 - 87 psi)